Inscriere cercetatori

Site nou !

Daca nu va puteti recupera parola (sau aveti alte probleme), scrieti-ne la pagina de contact. Situl vechi se gaseste la adresa old.ad-astra.ro

Facebook

The safety of environment in final disposal of Ultima Gold scintillation liquid cocktail used for determination of radioactive content in various samples at Cernavoda Nuclear Power Plant

Domenii publicaţii > Chimie + Tipuri publicaţii > Articol în revistã ştiinţificã

Autori: Magdalena DIANU, Corneliu Podinã

Editorial: Editura Academiei Române, Revue Roumaine de Chimie, 52 (5), p.509-519, 2007.

Rezumat:

The Ultima Gold scintillation liquid waste samples was counted in a 2100 TRI-CARB Packard model Liquid Scintillation Analyzer using the Ultima Gold XR scintillation liquid cocktail. To determine tritium content, three large sample/cocktail plastic vials were prepared and four sets of measurements were performed. Counting errors and statistics were applied on them to obtain specific activity of the waste. The immobilization and packaging of Ultima
Gold liquid scintillation waste were carried out by cementation, using Portland cement and an adequate solidification additive in a standard dimension container. Radioactive waste packages produced comply with waste acceptance criteria developed by the National Commission for Nuclear Activities Control for disposal at the Bãiţa National Repository, Bihor County, and thus the protection of human health and safety of the environment are ensured.

Cuvinte cheie: chemistry, chemical engineering

URL: http://www.ear.ro/